VulnTideExploited vulnerabilities, by vendor. Data as of .
Not in CISA KEVReported exploited: VulnCheck KEV since Published CVSS 7.7 highEPSS 1.1% probability of exploitation in 30 days
Contec Health

CVE-2025-0626: Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

Is it exploited?

Reported, not confirmed by CISA. VulnCheck KEV lists it as exploited in the wild. CISA has not added it to its Known Exploited Vulnerabilities catalog as of .

VulnCheck KEV lists it since . It cites 1 public report, the first from .

How likely is exploitation in the next 30 days?
1.1%

EPSS score, percentile 66.8 of all scored records. A probability estimated by FIRST, not a report of exploitation.

How severe?
7.7 High

CVSS v4.0 base score, given by the CNA

What should I do?

CISA has set no required action, because the vulnerability is not in KEV.

The CVE record links no vendor advisory, patch or mitigation.

Which versions are affected?

Affected

  • All versions

Versions by product

Source: CISA KEV 2026.10.29, as of ; VulnCheck KEV, as of ; FIRST EPSS, as of ; CVE List, as of . About the data.

Description

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Source: CVE List, as of . About the data.

What to do

Advisories and patches

The CVE record links no vendor advisory, patch or mitigation.

Workaround (from the CVE record)

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .

This summary is based on the CVE record as of . Check the vendor's advisory before you act: affected versions and fixes can change.

Affected products and versions

Versions by product, as listed in the CVE record (1 product)
Contec Health CMS8000 Patient Monitor
StatusVersions
AffectedAll versions

Versions not listed are: unaffected.

Source: CVE List, as of . About the data.

Weakness

CWE-912

CWE and CAPEC are maintained by MITRE; the entries are those named in the CVE record.