VulnTideExploited vulnerabilities, by vendor. Data as of .
Not in CISA KEVReported exploited: VulnCheck KEV since Published CVSS 8.2 highEPSS 0.8% probability of exploitation in 30 days
Contec Health

CVE-2025-0683: Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

Is it exploited?

Reported, not confirmed by CISA. VulnCheck KEV lists it as exploited in the wild. CISA has not added it to its Known Exploited Vulnerabilities catalog as of .

VulnCheck KEV lists it since . It cites 2 public reports, the first from .

How likely is exploitation in the next 30 days?
0.8%

EPSS score, percentile 56.2 of all scored records. A probability estimated by FIRST, not a report of exploitation.

How severe?
8.2 High

CVSS v4.0 base score, given by the CNA

What should I do?

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. Full textSolution from the CVE record.

CISA has set no required action, because the vulnerability is not in KEV.

The CVE record links no vendor advisory, patch or mitigation.

Which versions are affected?

Affected

  • All versions

Versions by product

Source: CISA KEV 2026.10.29, as of ; VulnCheck KEV, as of ; FIRST EPSS, as of ; CVE List, as of . About the data.

Description

Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

Source: CVE List, as of . About the data.

What to do

Advisories and patches

The CVE record links no vendor advisory, patch or mitigation.

Solution (from the CVE record)

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .

This summary is based on the CVE record as of . Check the vendor's advisory before you act: affected versions and fixes can change.

Affected products and versions

Versions by product, as listed in the CVE record (1 product)
Contec Health CMS8000 Patient Monitor
StatusVersions
AffectedAll versions

Versions not listed are: unaffected.

Source: CVE List, as of . About the data.

Weakness

CWE-359
  • CWE-359: Exposure of Private Personal Information to an Unauthorized Actor

CWE and CAPEC are maintained by MITRE; the entries are those named in the CVE record.

Timeline

DateEvent
CVE record published
CVE record last updated

References

2